Remote Commands
The Common Commands
Table 8-5 The Standard Event Status Register
Example  OUTPUT 728;"*ESR?"
ENTER 728; A$
*IDN? 
Syntax  *IDN?
Definition  The identification query commands the 
instrument to identify itself over the interface.
Response: HEWLETT-PACKARD, 
HP8156A, mmmmmmmmmm, n.nn
Example  DIM A$ [100]
OUTPUT 728;"*IDN?"
ENTER 728; A$
BITS MNEMONICS  BIT VALUE 
7 Power On 128
6User Request 64
5 Command Error 32
4 Execution Error 16
3 Device Dependent Error 8
2 Query Error 4
1 Request Control 2
0 Operation Control 1
HEWLETT-PACKARD: manufacturer
HP8156A: instrument model number
mmmmmmmmmm: serial number (not supplied)
n.nn: firmware revision level