user manual
 111
Remote Commands
OUTPut Commands
calibration factor (see ) to set the attenuation 
factor to the required value for use as the base 
value for the through-power
Cal
New
= (Through - Power
Base
- Att) + Cal
Current
When you switch the absolute power mode 
OFF, the last set calibration factor becomes 
active, and the attenuation factor is set so that 
the filter attenuation does not change. That is
Att(dB) = Attenuation
filter
(dB) + Cal(dB)
NOTE Using any of the :INPut:ATTenuation commands or queries, or 
any of the :INPut:OFFSet commands or queries, switches the 
absolute power mode off automatically.
See “:OUTPut:POWer” on page 112 for 
information on setting the through-power.
Switch the mode off (using OFF or 0) to set the 
attenuation of the filter by specifying the 
attenuation and calibration factors. Switch the 
mode on (using ON or 1) to set the attenuation 
of the filter by specifying the through-power.
:OUTPut:APMode? 
Syntax  :OUTPut:APMode?
Description  The query returns whether the attenuation of 
the filter is set by the attenuation and 
calibration factors, or by the through-power.
0 indicates the instrument sets the attenuation 
of the filter from the attenuation and calibration 
factors. 1 indicates that the instrument sets the 
attenuation of the filter from the through-
power.