User`s guide

512 VEE User’s Guide
Appendix
Figure 287 Mask Test, Step 2
Key Points
Mask: The mask is created using the Data Constant Coord
object, then configuring it for five array elements. You input the
coordinate pairs separated by commas and VEE adds the
parentheses. The x values were chosen knowing that the time
span of the waveform was 20 milliseconds. Also, note that the
Waveform (Time) display will accept a Coord data type as an
input. You could also use a Data Build Data Arb Waveform
object, which converts a Coord to a Waveform data type by
specifying the number of points in the Waveform.
Comparator: This object compares a test value against a
reference value. Once again, you can compare a waveform to an
array of coordinate pairs. The Failures pin gives you an array of
the data points that failed, which you can send to the display
and highlight with a different color or type of line.