Technical data
1 Calibration Procedures
Performance Test Record for U8032A
26 U8030A Series Service Guide
Performance Test Record for U8032A
Table 1 - 4 U8032A CV performance test record
Test description Actual result
Specifications
Upper limit Lower limit
CV programming accuracy @ 0 volts (DVM
reading
[1]
)
Output 1 +0.015 V –0.015 V
Output 2 +0.015 V –0.015 V
Output 3 +0.25 V –0.25 V
CV readback accuracy @ 0 volts
Output 1 DVM + 0.010 V DVM – 0.010 V
Output 2 DVM + 0.010 V DVM – 0.010 V
CV programming accuracy @ full scale (DVM
reading)
Output 1 60.165 V 59.835 V
Output 2 60.165 V 59.835 V
Output 3 5.25 V 4.75 V
CV readback accuracy @ full scale
Output 1 DVM + 0.165 V DVM – 0.165 V
Output 2 DVM + 0.165 V DVM – 0.165 V
CV load regulation Maximum change:
Output 1 <8 mV
Output 2 <8 mV
Output 3 ≤5 mV