Technical data

Calibration Procedures 1
Performance Verification Tests
U8030A Series Service Guide 15
6 Configure the differential amplifier as follows.
Enable the AC mode (positive and negative) to remove
the DC component
Enable the differential mode
Set the gain to ×10
Set the attenuation to 1
Set the low- pass filter to 20 MHz bandwidth limit to
filter out input signals that contain higher frequencies
Enable the zero precision voltage generator
Set the input impedance to 50 Ω
7 Configure the scope as follows.
Set the time range to 50 ms (20 Hz)
Acquire every single sample at the maximum sampling
rate and retain only the minimum and maximum values
in a “sampling region” (50 ms time range)
Enable the 20 MHz cut- off frequency for better high
frequency cut- off
Enable AC coupling
Enable auto- triggering
8 Allow the scope run for a few seconds to generate enough
measurement points.
9 Obtain the maximum peak- to- peak voltage measurement
as indicated in the scope. Divide this value by 10 to get
the CV peak- to- peak noise measurement. The result
should not exceed
10 mV
pp
for U8031A and U8032A, and
50 mV
pp
for Output 3.
10 Configure the RMS voltmeter as follows.
Set the high- pass filter to 10 Hz
Enable AC coupling