Technical data
Calibration Procedures 1
Test Considerations
U8030A Series Service Guide 5
Test Considerations
For optimum performance, all procedures should comply
with the following recommendations:
• Ensure that the calibration ambient temperature is stable
and is recommended at 23 °C ± 2 °C.
• Ensure that the ambient relative humidity is less than
80%.
• Allow a warm- up period of 1 hour by powering on the
power supply. During the warm- up period, set the power
supply to its maximum programmable voltage and current.
Then, enable the output.
• Keep the test setup connection cables as short as possible.