Technical data
Calibration Procedures 1
Recommended Test Equipment
U8030A Series Service Guide 3
Recommended Test Equipment
The test equipment recommended for the performance
verification and adjustment procedures is listed below in
Table 1- 1. If the exact instrument is not available, substitute
with another calibration standard of equivalent accuracy.
Table 1 - 1 Recommended test equipment
Equipment Requirement(s) Recommended model Purpose Used
[1]
Digital multimeter
(DMM)
8½ digits resolution Agilent 3458A Measures accurate DC voltage
PV
C
Oscilloscope
• Sensitivity: 1 mV
• Bandwidth limit: 20 MHz
• Probe:
• 1:1 with RF tip
• 10:1 with RF tip for > 50 V
Agilent DSO8064A or
Infiniium equivalent
Displays transient response.
Displays ripple and noise
waveform.
PV
Differential
amplifier
• Bandwidth > 20 MHz
(bandwidth limit: 20 MHz)
• AC coupling
• Amplifier × 10
LeCroy DA1855A Signal conditioning preamplifiers PV
RMS volt meter 20 Hz to 20 MHz R&S URE3 Measures RMS noise PV
50 Ω feed-through
termination
- Pomona
Loads termination to match
output impedance of the
differential amplifier
PV
AC/DC current
converter
-
Tektronix TCP305 and
Tek t r on i x T C PA3 0 0
Converts current to voltage PV
Electronic load
• Voltage range: 60 Vdc
• Current range: 6 A
• Open and Short switches
• Transient O n/Off
Agilent 6060B
Measures load and line
regulations and transient
response time
PV
AC power source
Capable of supplying 90 Vac to
250 Vac
Agilent 6813B
Functions as variable voltage
transformer
PV