Technical data
88 U4154A Logic Analyzer Service Guide
6 Returning and Replacing a U4154A Module or its cables
To return the U4154A module or cable for Repair/Exchange
Before returning the U4154A module or a logic analyzer
cable to Agilent Technologies, contact your nearest Agilent
Technologies Sales Office for additional details. Information
on contacting Agilent can be found at www.agilent.com.
1 Write the following information on a tag and attach it to
the module/cable.
• Name and address of owner.
• Model number.
• Serial number.
• Description of service required or failure indications.
2 Remove accessories from the module.
Only return accessories to Agilent Technologies if they are
associated with the failure symptoms.
3 Package the module/cable.
You can use either the original shipping containers, or
order materials from an Agilent Technologies sales office.
4 Seal the shipping container securely, and mark it
FRAGILE.
CAUTION
For protection against electrostatic discharge (ESD), package the
module in ESD-safe material.