Technical data

85
Agilent U4154A Logic Analyzer
Service Guide
Agilent Technologies
6
Returning and Replacing a U4154A
Module or its cables
To remove the U4154A module 86
To remove the logic analyzer cable 86
To install the U4201A logic analyzer cable 87
To replace the circuit board 87
To return the U4154A module or cable for Repair/Exchange 88
This chapter contains the instructions for removing and
replacing the U4154A logic analyzer module, and the probe
cables of the module as well as the instructions for
returning defective parts to Agilent Technologies.
CAUTION
Turn off the AXIe chassis before installing, removing, or replacing a
module in the chassis.
The enclosure surface of the U4154A module may become hot
during use. If you need to remove the module, first power down the
AXIe chassis, allow the module to cool, and then pull the module out
of the chassis.
CAUTION
Electrostatic discharge can damage electronic components. Use
grounded wriststraps and mats when performing any service to this
module.