Technical data
Troubleshooting 5
U4154A Logic Analyzer Service Guide 77
2 Set up the stimulus board
a Configure the oscillator select switch S1 according to
the following settings:
• S1 0 (Off).
• S2 1 (On).
• S3 0 (Off).
• Int.
b Configure the data mode switch S4 according to the
following settings:
• Even.
• Count.
c Press the Resynch VCO button, then the Counter RST
(Counter Reset) button.
3 In the Agilent Logic Analyzer application, choose
File→New. This puts the logic analysis system into its
initial state.