Technical data

Troubleshooting 5
U4154A Logic Analyzer Service Guide 77
2 Set up the stimulus board
a Configure the oscillator select switch S1 according to
the following settings:
S1 0 (Off).
S2 1 (On).
S3 0 (Off).
Int.
b Configure the data mode switch S4 according to the
following settings:
Even.
Count.
c Press the Resynch VCO button, then the Counter RST
(Counter Reset) button.
3 In the Agilent Logic Analyzer application, choose
FileNew. This puts the logic analysis system into its
initial state.