Technical data
62 U4154A Logic Analyzer Service Guide
3 Testing U4154A Performance
Performance Test Record
LOGIC ANALYZER MODEL NO. (circle one): U4154A
Logic Analyzer Serial No. Work Order No.
Date: Recommended Test Interval - 2 Year/4000 hours
Recommended next testing:
TEST EQUIPMENT USED
Pulse Generator Model No. Oscilloscope Model No.
Pulse Generator Serial No. Oscilloscope Serial No.
Pulse Generator Calibration Due Date: Oscilloscope Calibration Due Date:
MEASUREMENT UNCERTAINTY
Clock Rate
Pulse Generator Frequency Accuracy:
81134A: ±0.005% of setting.
8133A: ±0.5% of setting.
2% = uncertainty + at least 1% test margin.
Setting
1250 MHz + 2% = 1275 MHz
TEST RESULTS
Logic Analysis System Self-Tests (Pass/Fail):
Performance Test: Maximum State Data Rate
Pulse Generator Settings Freq: 1250 MHz plus 2% (1275 MHz).
Square wave.
Pulse Generator Frequency Test (Pass/Fail)
Final Pulse Generator Frequency
Test Criteria Te s t 1 o f 2
Eye Scan locates an eye for each bit
Test 2 of 2
Correct number of occurrences
detected
Normal Complement Normal Complement
Pod 1 Results (pass/fail):
Pod 2 Results (pass/fail):