Technical data
Testing U4154A Performance 3
U4154A Logic Analyzer Service Guide 61
The activity indicator will show activity on clock 1.
3 Set the State Clock options:
a For State Clock, select Pod 1 Clock and Both edges.
4 Perform the normal and complement tests for Pod 3,
changing the connection to the pod, setting thresholds,
making channel assignments, etc., as appropriate.
Repeat the above steps for testing pods 4 to 8. Upon
completion, the logic analyzer is completely tested.
Conclude the State Mode Tests
Do the following steps to properly shut down the logic
analyzer session after completing the state mode tests.
1 End the test.
a From the Main Menu, choose File→Exit. At the dialog
“Do you want to save the current configuration?” select
No.
Ending and restarting the logic analysis session will
re- initialize the system.
b Disconnect all cables and adapters from the pulse
generator and the oscilloscope.