Technical data
56 U4154A Logic Analyzer Service Guide
3 Testing U4154A Performance
4 From the Main Menu, choose MarkersβNew.
a You can accept the default name for the new marker.
b Change the Position field to Va lue .
c Select the Occurs... button and create the marker setup
shown below.
5 In the Value window, select the Properties... button.
6 In the Value Properties window, select Stop repetitive run
when value is not found.
7 Select OK to close the marker Value Properties window.
8 Select OK to close the marker Value window. The system
will search the display for the occurrences specified.
9 Select OK to close the New Marker window.