Technical data
Testing U4154A Performance 3
U4154A Logic Analyzer Service Guide 55
Close the Eye Scan and Analyzer Setup Windows
1 Select OK to close the Eye Scan - Sample Positions and
Threshold Settings dialog.
2 Select OK to close the Analyzer Setup window.
Configure the markers
Data must be acquired before the markers can be configured.
Therefore, you need to run the analyzer to acquire data.
1 Switch to the Listing window by selecting the Listing tab
at the bottom of the main window.
2 Select the Run icon .
3 Data will appear in the Listing Window upon completion
of the run.
If the data values are not “A”s and “5”s, you may need to
set the sampling positions in different eyes.