Technical data

Testing U4154A Performance 3
U4154A Logic Analyzer Service Guide 33
3 Set up the oscilloscope.
a Set up the oscilloscope according to one of the
following tables.
Table 7 DSO90604A Oscilloscope Setup
Setup: Channel 1 Setup: Channel 2 Setup: Channel 3 Setup: Channel 4
On On Off Off
Scale: 50 mV/div Scale: 50 mV/div
Offset: 600 mV Offset: 600 mV
Skew: (Set later. See
page 39.)
Skew: 0.0 seconds
Setup: Horizontal Setup: Trigger Setup: Acquisition Setup: Display
Scale: 200 ps Mode: Edge Sampling Mode: Real Time,
Normal
Waveforms
Position: 0 s Source: Channel 1 Memory Depth: Automatic Connect dots
Reference: Center Level: 600 mV Averaging: Enabled Color Grade: not selected
External 10 MHz Reference
Clock: not selected
Edge: Rising Edge # of Averages: 16 Infinite Persistence: not
selected
Zoom: not selected Sweep: Auto Sampling Rate: Automatic Waveform Brightness: as
preferred
Sine(x)/x Interpolation:
checked
Grid: On, Quantity: 1
Intensity: as preferred
Measure: Markers
Mode: Manual placement
All else: (n/a)