Technical data
Testing U4154A Performance 3
U4154A Logic Analyzer Service Guide 25
The progress of the self tests is displayed in the Results
area of the window.
5 When the self- tests are complete, check the Results
window to ensure that the Result Summary says that all
tests passed. If all tests did not pass, refer to “To use the
flowcharts" on page 68.
6 Select the Close button to close the Analysis System Self
Tests window.
7 If all module self-tests pass, then record “PASS” in the
“Logic Analysis System Self- Tests” section of the
Performance Test Record (page 62).
Equipment Required for the Performance Test
The following equipment is required for the performance test
procedure.
Table 3 Equipment Required
Equipment Critical Specification Recommended Model/Part
Pulse Generator ≥ 1275 MHz, two channels, differential
outputs, 150-180 ps rise/fall time (if
faster, use transition time converters)
Agilent 81134A or equivalent
150 ps Transition Time Converter
(Qty 4)
Required if pulse generator’s rise time
is less than 150 ps. (Pulse generator
conditions: Voffset=1V, ΔV=250 mV.)
Required for 81134A or 8133A opt. 003.
Agilent or HP 15435A
Oscilloscope Bandwidth ≥ 6 GHz Agilent DSO90604A or equivalent
SMA Coax Cable
(Qty 2)
>18 GHz bandwidth Agilent or HP 8120-4948
Flying Lead Probe Set
(Qty 2)
no substitute Agilent E5382A
Male BNC to Female SMA adapters
(Qty 2)
Cambridge Products CP-AD507
(see www.cambridgeproducts.com)
SMA/Flying Lead test connectors, (f)
SMA to (f) SMA to Flying Lead Probe
(Qty 4)
no substitute See “Assemble the SMA/Flying Lead
Tes t C o n ne ct o rs " on page 26