Technical data

24 U4154A Logic Analyzer Service Guide
3 Testing U4154A Performance
Perform the Self-Tests
Once you have connected all the hardware components for
the U4154A module and created a logical module for U4154A
in the Agilent Logic Analyzer application, you are ready to
run the self tests on U4154A.
1 Before performing the self- tests, disconnect all probes
from the logic analyzer module.
2 Select Help->Self-Test... from the main menu. The Analysis
System Self Tests window will appear.
3 In the Select suites list, select U4154A. Then, select All in
the Select tests list.
4 Select Start. This will perform a complete system self- test.