Technical data
21
Agilent U4154A Logic Analyzer
Service Guide
Agilent Technologies
3
Testing U4154A Performance
Perform the Self-Tests 24
Equipment Required for the Performance Test 25
Assemble the SMA/Flying Lead Test Connectors 26
Set Up the Test Equipment 31
Connect the Test Equipment 35
Test the U4154A Module 40
Performance Test Record 62
This chapter tells you how to test the performance of the
U4154A logic analyzer against the specifications listed on
page 13.
To ensure the U4154A logic analyzer module is operating as
specified, software tests (self- tests) and a manual
performance test are done. The logic analyzer is considered
performance- verified if all of the software tests and the
manual performance test have passed.
The specifications for the U4154A logic analyzer define a
maximum state data rate at which data can be acquired in
state mode. The manual performance test (maximum state
data rate test) verifies that the logic analyzer meets these
specifications.
AXIe Chassis
You must test the U4154A logic analyzer module in a
M9502A or M9505A AXIe chassis.
Test Strategy
Only specified parameters are tested. Specifications are
listed on page 13. The test conditions defined in this
procedure ensure that the specified parameter is as good as
or better than specifications. No attempt is made to
determine performance which is better than specifications.
Not all channels of the logic analyzer will be tested; a