Technical data

17
Agilent U4154A Logic Analyzer
Service Guide
Agilent Technologies
2
Preparing for Troubleshooting or
Performance Testing
To set up the U4154A module 18
To test the U4154A module 19
To clean t h e mod u l e 19
This chapter gives you instructions for preparing the U4154A
logic analyzer module for troubleshooting or servicing it.
Operating Environment
The operating environment is listed on page 15. Note the
non- condensing humidity limitation. Condensation within the
instrument can cause poor operation or malfunction. Provide
protection against internal condensation.
The logic analyzer module will operate at all specifications
within the temperature and humidity range given on
page 15. However, reliability is enhanced when operating the
module within the following ranges:
Temperature: +20°C to +35°C (+68°F to +95°F)
Humidity: 20% to 80% non- condensing