User`s guide
2 Operation and Features
Math Operations
70 U3606B User’s Guide
Limit
The limit test function enables you to perform pass or fail
testing to the upper and lower limits that you specify. You
can set the upper and lower limits to any value between 0
and ±120% of the highest range for the present measurement
function. The upper limit value you select must be larger
than the lower limit value. The initial factory settings for
both values is zero.
NOTE
Each time a new minimum, maximum, or average value is stored, the
instrument beeps once (if the beeper is enabled). The U3606B calculates
the average of all readings and records the number of readings taken since
the MinMax function was enabled.
Accumulated statistics are:
• Max: maximum reading since the MinMax function was enabled
• Min: minimum reading since the MinMax function was enabled
• Avg: average of all readings since the MinMax function was enabled
• MaxMinAvg: present reading (actual input signal value)
NOTE
• This math operation applies to all measurement functions except
continuity and diode test.
• The instrument clears all limits after a Factory Reset (
*RST command),
an Instrument Preset (SYSTem:PRESet command), or when a
measurement function is changed.
U3606-90054_UG_EN.book Page 70 Friday, July 11, 2014 5:12 PM