User`s guide

2 Operation and Features
Measuring Frequency/Pulse Width/Duty Cycle (Voltage Path)
54 U3606B User’s Guide
Probe the test points and read the display
In this example, we obtain a frequency measurement of 12.345 kHz,
with autoranging enabled and 5½ digit resolution selected by default.
5
CAUTION
If the frequency signal measured is below 20 Hz, you must manually
set the range of the AC voltage or AC current measurement to acquire
a stable reading.
NOTE
The range and resolution of the frequency, pulse width, and duty cycle
measurement follows the configuration of the AC voltage or AC current
measurement (dependent on the path chosen).
The AC voltage or AC current measurement display will flash briefly
before the frequency measurement display is shown.