User`s guide
2 Operation and Features
Measuring Current
34 U3606B User’s Guide
Current measurement summary
Probe the test points and read the display
In this example, we obtain an AC current measurement of 1.2345 A,
with autoranging enabled and 5½ digit resolution selected by default.
5
Table 2 - 2 Current measurement summary
Item Description
DC current measurement
Available ranges 10.0000 mA, 100.000 mA, 1.00000 A, 3.0000 A
Measurement
method
Sigma Delta A-to-D converter
Burden voltage
and shunt
resistance
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.05 Ω for 1 A range
• < 0.7 V, 0.05 Ω for 3 A range
Input protection Protected with a 3.15 A/500 V, FF fuse
U3606-90054_UG_EN.book Page 34 Friday, July 11, 2014 5:12 PM