User`s guide
Operation and Features 2
Measuring Voltage
U3606B User’s Guide 31
Voltage measurement summary
Probe the test points and read the display
In this example, we obtain a DC voltage measurement of 1.2345 V,
with autoranging enabled and 5½ digit resolution selected by default.
5
Table 2 - 1 Voltage measurement summary
Item Description
DC voltage measurement
Available ranges
19.9999 mV, 100.000 mV, 1.00000 V, 10.0000 V, 100.000 V,
1000.00 V
Measurement
method
Sigma Delta A-to-D converter
Input impedance
10 MΩ ± 2% range (typical) in parallel with capacitance
< 120 pF
Input protection 1000 V
rms
on all ranges
U3606-90054_UG_EN.book Page 31 Friday, July 11, 2014 5:12 PM