User`s guide
Characteristics and Specifications 5
DC Power Supply Specifications
U3606B User’s Guide 179
Sweep characteristics
Table 5-11 Scan output characteristics
Scan Constant voltage Constant current
Range S1 S2 S2m S1 S1m S2
Maximum
amplitude
[1]
31.500 V 8.400 V 1050.0 mV 1.0500 A 105.00 mA 3.1500 A
Step 1 step to 100 steps 1 step to 100 steps
Dwelling time 1 s to 99 s 1 s to 99 s
[1] Amplitude start position is fixed at 0 (V or A) by default.
Table 5-12 Ramp output characteristics
Ramp Constant voltage Constant current
Range S1 S2 S2m S1 S1m S2
Maximum
amplitude
[1]
31.500 V 8.400 V 1050.0 mV 1.0500 A 105.00 mA 3.1500 A
Step 1 step to 10000 steps 1 step to 10000 steps
Dwelling time 100 ms (typical) per step 100 ms (typical) per step
[1] Amplitude start position is fixed at 0 (V or A) by default.
U3606-90054_UG_EN.book Page 179 Friday, July 11, 2014 5:12 PM