User`s guide

5 Characteristics and Specifications
DC Power Supply Specifications
178 U3606B User’s Guide
[2] For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the duty cycle should be
calculated as:
Calculation example: Frequency setting = 4800 Hz, Duty cycle setting = 50%
The duty cycle accuracy (for frequency setting 4800 Hz) is calculated as 50% ± 5.1%.
[3] Characteristic applies when the positive or negative pulse width is greater than 50 μs.
[4] For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the pulse width should
be calculated as:
Calculation example: Frequency setting = 4800 Hz, Duty cycle setting = 50%
The pulse width accuracy (for frequency setting 4800 Hz and duty cycle setting 50%) is calculated as 0.1042 ms ±10.625 μs.
Accuracy 0.4%
frequency
100
----------------------------- 1
⎝⎠
⎛⎞
0.1%×+
⎝⎠
⎛⎞
=
Charateristics of duty cycle 0.4%
4800
100
------------ 1
⎝⎠
⎛⎞
0.1%×+±
5.1
100
---------± 5.1
± % ===
Accuracy
0.4%
frequency
100
----------------------------- 1
⎝⎠
⎛⎞
0.1%×+
⎝⎠
⎛⎞
frequency
--------------------------------------------------------------------------------
=
Characteristics of pulse width 0.4%
4800
100
------------ 1
⎝⎠
⎛⎞
0.1%×+
⎝⎠
⎛⎞
1
4800
------------×
⎝⎠
⎛⎞
±
5.1
100
---------
1
4800
------------×
⎝⎠
⎛⎞
± 10.625± μs===
NOTE
The rise and fall time are 25 μs typically between 10% and 90% of the
signal amplitude.
The additional load regulation is 0.15 V/A.
U3606-90054_UG_EN.book Page 178 Friday, July 11, 2014 5:12 PM