User`s guide

Characteristics and Specifications 5
DC Power Supply Specifications
U3606B User’s Guide 175
Programming resolution
Voltage 1 mV 1 mV 1 mV 1 mV 0.1 mV
Current 0.1 mA 0.1 mA 0.01 mA 0.1 mA 0.1 mA
Readback resolution
Voltage 1 mV 1 mV 1 mV 1 mV 0.1 mV
Current 0.1 mA 0.1 mA 0.01 mA 0.1 mA 0.1 mA
Front panel resolution
Voltage 1 mV 1 mV 1 mV 1 mV 0.1 mV
Current 0.1 mA 0.1 mA 0.01 mA 0.1 mA 0.1 mA
Transient response time
Less than 100 μs for output to recover to within 15 mV
following a change in output current from full load to
half load or vice versa.
Command processing time
Average time for output voltage to begin to change after
receipt of digital data when instrument is connected
directly to the USB or GPIB is less than 100 ms.
Over-voltage protection (for CC mode)
Accuracy: 0.5% + 0.5 V
Activation time
[2]
: < 2ms
Over-current protection (for CV mode)
Accuracy: 0.5% + 0.05 A
Activation time
[2]
: < 2 ms
[1] The terminal sense is related to the resistance of the contacts or leads, and proportional to the load condition.
[2] Average time for the detection of OVP or OCP condition. The output will be dropped down and set to standby within 20 ms.
Table 5 - 8 DC power supply performance specifications (continued)
Parameter
Specifications
S1S2 S1 S1m S2 S2m
Output ratings AUTO 30 V/1 A
100 mA
/30 V
8 V/3 A
1000 mV
/3 A
U3606-90054_UG_EN.book Page 175 Friday, July 11, 2014 5:12 PM