5 Characteristics and Specifications
Digital Multimeter Specifications
170 U3606B User’s Guide
Input protection: 750 V
rms
on all ranges
Crest factor: For < 5:1 errors included. Limited by the peak input
and 100 kHz bandwidth. Maximum 3.0 at full scale.
Peak input: 300% of range. Limited by maximum input.
Response time: Approximately 2.5 s when the displayed reading
reaches 99.9% AC rms value of the tested input signal
at the same range.
Overload ranging: Will select higher range if peak input overload is
detected during auto range. Overload is reported in
manual ranging.
AC current
Measurement method: AC coupled true rms
Maximum input current: 10 mA to 3.0 A DC or AC rms
[1]
Burden voltage and
shunt resistance:
• < 0.2 V, 10 Ω for 10 mA range
• < 0.2 V, 1 Ω for 100 mA range
• < 0.3 V, 0.05 Ω for 1 A range
• < 0.7 V, 0.05 Ω for 3 A range
Input protection: Protected with a 3.15 A/500 V, FF fuse
Crest factor: For < 5:1 errors included. Limited by the peak input
and 100 kHz bandwidth. Maximum 3.0 at full scale.
Peak input: 300% of range. Limited by maximum input.
Response time: Approximately 2.5 s when the displayed reading
reaches 99.9% AC rms value of the tested input signal
at the same range.
[1] Any current measurement greater than 500 mA will have a temporary
thermo-effect. If you wish to measure a smaller current or offset current
measurement immediately after a high current measurement, ensure that the
U3606B is cooled down.