User`s guide
5 Characteristics and Specifications
Digital Multimeter Specifications
166 U3606B User’s Guide
Table 5 - 4 Frequency sensitivity for voltage measurement
Input range
[1]
Minimum sensitivity (rms sine wave)
20 Hz to 100 kHz 100 kHz to 300 kHz 300 kHz to 1 MHz
100 mV 50 mV 50 mV 0.5 V
1.0 V 100 mV 120 mV 0.5 V
10 V 1 V 1.2 V -
100 V 10 V 12 V -
750 V 100 V - -
[1] Maximum input for specified accuracy = 10 × range or 750 V
rms
or 1000 V
dc
Table 5 - 5 Frequency sensitivity for current measurement
Input range Minimum sensitivity (rms sine wave)
20 Hz to 10 kHz
10 mA 1 mA
100 mA 10 mA
1.000 A 100 mA
3 A 300 mA
U3606-90054_UG_EN.book Page 166 Friday, July 11, 2014 5:12 PM