User`s guide
5 Characteristics and Specifications
Digital Multimeter Specifications
160 U3606B User’s Guide
Resistance
[4]
100.000 Ω 1 mΩ 0.83 mA 0.04 + 0.008 0.04 + 0.008 0.05 + 0.008
0.0050
+ 0.0005
1000.00 Ω 10 mΩ 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005
0.0050
+ 0.0005
10.0000 kΩ 100 mΩ 100 µA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005
0.0050
+ 0.0005
100.000 kΩ 1 Ω 10 µA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005
0.0050
+ 0.0005
1.00000 MΩ 10 Ω 900 nA 0.05 + 0.005 0.05 + 0.005 0.06 + 0.005
0.0050
+ 0.0005
10.0000 MΩ 100 Ω 205 nA 0.20 + 0.005 0.20 + 0.005 0.25 + 0.005
0.0150
+ 0.0005
100.000 MΩ 1 kΩ
205 nA ||
10 MΩ
1.60 + 0.005 1.60 + 0.005 2.00 + 0.005
0.1500
+ 0.0005
Low-
resistance
[5]
100 mΩ
0.01/
0.001 mΩ
1.0000 A - - 0.25 + 0.05 -
1000 mΩ 0.1/0.01 mΩ 100.00 mA - - 0.25 + 0.03 -
10 Ω 1/0.1 mΩ 100.00 mA - - 0.09 + 0.03 -
100 Ω 10/1 mΩ 10.00 mA - - 0.09 + 0.03 -
1000 Ω 0.1/10 mΩ 10.00 mA - - 0.09 + 0.03 -
Continuity 1.0000 kΩ 100 mΩ 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005
0.0050
+ 0.0005
Diode
[6]
1.0000 V 0.0001 V 0.83 mA 0.04 + 0.005 0.04 + 0.005 0.05 + 0.005
0.0050
+ 0.0005
Table 5 - 1 DC accuracy specifications ± (% of reading + % of range) (continued)
Function Range
[1]
Resolution Test current
or burden
voltage
24 Hours
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
U3606-90054_UG_EN.book Page 160 Friday, July 11, 2014 5:12 PM