User`s guide
Characteristics and Specifications 5
Digital Multimeter Specifications
U3606B User’s Guide 159
DC specifications
Table 5 - 1 DC accuracy specifications ± (% of reading + % of range)
Function Range
[1]
Resolution Test current
or burden
voltage
24 Hours
[2]
23 ºC ± 1 ºC
90 Days
23 ºC ± 5 ºC
1 Year
23 ºC ± 5 ºC
Temperature
coefficient
0 ºC to 18 ºC
28 ºC to 55 ºC
DC voltage
19.9999 mV 0.1 μV - 0.012 + 0.04 0.015 + 0.04 0.025 + 0.04
0.0015
+ 0.0040
100.000 mV 1 μV - 0.012 + 0.008 0.015 + 0.008 0.025 + 0.008
0.0015
+ 0.0008
1.00000 V 10 μV - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005
0.0010
+ 0.0005
10.0000 V 100 μV - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005
0.0020
+ 0.0005
100.000 V 1 mV - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005
0.0015
+ 0.0005
1000.00 V 10 mV - 0.012 + 0.005 0.015 + 0.005 0.025 + 0.005
0.0015
+ 0.0005
DC current
[3]
10.0000 mA 0.1 μA < 0.2 V 0.05 + 0.015 0.05 + 0.015 0.05 + 0.015
0.0060
+ 0.0005
100.000 mA 1 μA < 0.2 V 0.05 + 0.005 0.05 + 0.005 0.05 + 0.005
0.0060
+ 0.0005
1.00000 A 10 μA < 0.3 V 0.05 + 0.007 0.05 + 0.007 0.15 + 0.007
0.0100
+ 0.0005
3.0000 A 100 μA < 0.7 V 0.05 + 0.007 0.05 + 0.007 0.15 + 0.007
0.0150
+ 0.0010
U3606-90054_UG_EN.book Page 159 Friday, July 11, 2014 5:12 PM