User`s guide
System-Related Operation 4
Utility Menu Items
U3606B User’s Guide 145
Changing the power-on state
The following procedure shows you how to enable or disable
the automatic recall of the power- off state when the power
is turned on. For remote interface operation, refer to the
MEMory:STATe:RECall:AUTO command in the U3606B
Programmer’s Reference.
Press [] or [] to change the power- on state.
• Select “rESEt” to automatically reset the instrument to the
factory default state when the power is turned on.
• Select “LASt” to automatically recall the last power- off
state of the instrument when the power is turned on.
NOTE
The power-off state includes
• the multimeter function, autoranging state, measurement range and
resolution, and frequency path
• the DC power supply function, range, remote sensing, square-wave,
amplitude, frequency, duty-cycle, and pulse-width
The power-off state does not include
• math operations
• ramp, scan, and protection and limitation
[1]
functions
[1] Although the U3606B does not store the power-off protection and limitation of the source, it can recall your
saved settings by remote or local operation when you select “LASt”.