User`s guide

System-Related Operation 4
Utility Menu Items
U3606B User’s Guide 137
Press [] or [] to change the scan signal amplitude end
position.
Press [] or [] until the cursor is position on the scan
signal number of steps. Press [] or [] to change the
number of steps for the scan signal to increment from zero
to the amplitude end position.
Press [] or [] until the cursor is position on the scan
signal dwelling time. Press [] or [] to change the scan
signal dwelling time.
CV/CC scan signal parameters Scan signal range
Amplitude end position
(NN.NNN V or N.NNNN A)
Number of
steps (NNN)
Scan dwelling
time (NN s)
NOTE
The increment of each step in the scan signal will be the amplitude end
position divided by the number of steps. For example, a 15 V amplitude
end position divided by 100 steps gives an increment of 0.15 V per step.
The scan signal will “dwell” in the present step for the length of time
stated in the scan dwelling time before incrementing to the next step.
U3606-90054_UG_EN.book Page 137 Friday, July 11, 2014 5:12 PM