User`s guide

4 System-Related Operation
Utility Menu Items
136 U3606B User’s Guide
Configuring the scan signal parameters
Press [Voltage] or [Current] if you wish to configure the scan
signal parameters for CV or CC output.
Select an appropriate output range by pressing [Shift] >
[Range]. The scan signal amplitude end position is limited by
the range selected.
Table 4 - 2 Scan signal parameters
Scan signal Output
Items Constant Voltage Constant Current
Range
[1]
S1 S2 S2m S1 S2 S1m
Amplitude end position
[2]
0 V to
31.500 V
0 V to
8.400 V
0 V to
1050.0 mV
0 A to
1.0500 A
0 A to
3.1500 A
0 A to
105.00 mA
Number of steps 1 step to 100 steps
Dwelling time 1 s to 99 s
[1] Range S1S2 (autoranging) is not supported for the Scan function.
[2] Amplitude start position is fixed at 0 (V or A) by default.
U3606-90054_UG_EN.book Page 136 Friday, July 11, 2014 5:12 PM