3 DC Power Supply Operation
Remote Sensing
122 U3606B User’s Guide
With remote sensing, any voltage dropped in the load leads
must be added to the load voltage to calculate maximum
output voltage. The performance specifications are not
guaranteed when the maximum output voltage is exceeded.
Output noise
Any noise picked up on the sense leads also appears at the
output of the U3606B and may adversely affect the voltage
load regulation. Twist the sense leads to minimize external
noise pickup and run them parallel and close to the load
leads. In noisy environments it may be necessary to shield
the sense leads. Ground the shield at the U3606B end only.
Do not use the shield as one of the sense conductors.