Remote Sensing
120 U3606B User’s Guide
Enable remote sensing
To configure the U3606B for remote sensing:
1 Turn off the U3606B.
2 Remove the connections between the U3606B sensing (S+
and S–) and output (+ and –) terminals. Using a shielded
two- wire cable, connect the U3606B sensing terminals to
the load as shown in Figure 3- 1. Observe the polarity
when connecting the sensing leads to the load.
3 Turn on the U3606B.
4 Press [Voltage] to select the constant voltage mode. Use
the arrow keys to select the appropriate constant voltage
value.
5 Press [Shift] > [EXT] to enable remote sensing. When the
U3606B is operating in remote sensing mode, the EXT
annunciator on the front panel is illuminated.
6 Press [] to regulate the output voltage.
7 Press [Shift] > [EXT] again to disable remote sensing when
the output state is on standby.