User`s guide
3 DC Power Supply Operation
Remote Sensing
118 U3606B User’s Guide
4 Loosen the top screws with a slotted screw driver and
connect the rear output terminal block sensing (S+ and
S–) and output (+ and –) terminals to the load as shown
in Figure 3- 1 using a shielded two- wire cable.
Do not use the shield as one of the sensing conductors.
Ground the shield at the U3606B end only. The other end
of the shield should be left unconnected. Observe the
polarity when connecting the sensing leads to the load.
You can connect the output leads to either one of the two
+ or – terminals. They are internally shorted.
5 Tighten the top screws of the rear output terminal block
to secure firmly the sensing and output connections.
6 Place the rear output terminal block back in place and
tighten the two captive screws again.
NOTE
The sensing outputs (S+ and S–) must not be left unconnected. It must be
connected locally (Figure 3-2) or remotely (Figure 3-1).