User`s guide

DC Power Supply Operation 3
Selecting a Range
U3606B User’s Guide 113
Selecting a Range
You can allow the U3606B to automatically select the range
using autoranging or you can select a fixed range using
manual ranging.
Key Description
Press [Shift] > [Range] to toggle between the available ranges.
Table 3 - 6 Available ranges for DC power supply functions
Range CV
[1]
[1] Available ranges affect sweep functions (ramp and scan), OCP, and OC values.
CC
[2]
[2] Available ranges affect sweep functions (ramp and scan), OVP, and OV values.
Square wave
S1 (30 V/1 A)
✔✔✔
S2 (8 V/3 A)
✔✔✔
S1m (30 V/100 mA) -
-
S2m (1000 mV/3 A)
--
S1S2 (autoranging)
✔✔✔
NOTE
The S1 range is selected by default. You cannot change the range when
the output is enabled (OUT). The output should always be on stand-by
(SBY) before the range or output function can be changed.
The protection and limit values will always be set to the previous saved
value with respect to the range selected when the range is changed.
U3606-90054_UG_EN.book Page 113 Friday, July 11, 2014 5:12 PM