User`s guide

DC Power Supply Operation 3
Protection Functions
U3606B User’s Guide 99
NOTE
After adjusting the OC value, you can also press [Limit] again to save
the changes made or press [Shift] > [Exit] to discard the changes
made.
The OC feature is always enabled by default whenever you select
constant current mode. You cannot disable the OC feature.
If the OC value is set to a greater value than the OCP value, the OCP
value will be adjusted to equal the OC value.
If the OC value is set to zero, this will result in the output voltage being
dropped to zero for limiting.
The OC value is not applicable for square-wave output.
The OC value is limited by the range selected. Press [Shift] > [Range]
to select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).
Table 3 - 4 CV mode over-current range and values
Range
Over-current value
Minimum Maximum
S1 (30 V/1 A) 0 A 1.05 A
S2 (8 V/3 A) 0 A 3.15 A
S2m (1000 mV/3 A) 0 A 3.15 A
S1S2 (autoranging) N/A
[1]
[1] If the S1S2 (autoranging) range is selected, you will not be able to adjust the OC
value.
U3606-90054_UG_EN.book Page 99 Friday, July 11, 2014 5:12 PM