Protection Functions
96 U3606B User’s Guide
NOTE
After adjusting the OV value, you can also press [Limit] again to save
the changes made or press [Shift] > [Exit] to discard the changes
made.
The OV feature is always enabled by default whenever you select
constant current mode. You cannot disable the OV feature.
If the OV value is set to a greater value than the OVP value, the OVP
value will be adjusted to equal the OV value.
If the OV value is set to zero, this will result in the output current being
dropped to zero for limiting.
The OV value is not applicable for square-wave output.
The OV value is limited by the range selected. Press [Shift] > [Range]
to select an appropriate range. You can only select a range when the
output is disabled (the SBY annunciator is illuminated).
Table 3 - 3 CC mode over-voltage range and values
Range
Over-voltage value
Minimum Maximum
S1 (30 V/1 A) 0 V 31.5 V
S1m (30 V/100 mA) 0 V 31.5 V
S2 (8 V/3 A) 0 V 8.4 V
S1S2 (autoranging) N/A
[1]
[1] If the S1S2 (autoranging) range is selected, you will not be able to adjust the OV
value.