3 DC Power Supply Operation
Protection Functions
90 U3606B User’s Guide
Over-current protection (OCP)
In constant voltage mode, the U3606B regulates the output
voltage at the selected value, while the load current varies
as required by the load. The over- current protection will
disable the output if the load effect exceeds the programmed
protection value. This protection is useful when the load is
sensitive to an over- current condition.
The following steps shows you how to set the OCP trip level,
how to check OCP operation, and how to clear the OCP
condition.
NOTE
Setting the OCP trip level will not activate the OCP feature. To activate the
OCP feature, you will need to enable the output protection state from the
Utility menu.
Select the constant voltage function
1
The CV annunciator will turn on according to your selection.
2