Technical data

6 Specifications and Characteristics
104 U3402A User’s and Service Guide
Gate time
0.1 second or 1 period of the input signal, whichever is longer
Input impedance
1 MW ± 2% in parallel with <120 pF of all ranges
Maximum input voltage
750 V RMS /1100 V PEAK
2x10
7
V-Hz product on any range, normal mode input
1x10
6
V-Hz product on any range, common mode input
Input protection
750 V RMS on all ranges
Response time
Approximately 1.5 seconds when the displayed reading reaches 99.9% of frequency value
Measurement Noise Rejection
Common mode reject ratio (CMRR) for 1 kW unbalanced LO lead
50/60 Hz ± 0.1%: DC > 90 dB
Normal mode rejection ratio (NMRR)
50/60 Hz ± 0.1%: > 50 dB
dBm Operation
0 dBm
1 mW at 600 W reference impedance
Resolution
Slow: 0.01 dB for all ranges
Medium: 0.01 dB for all ranges
Fast: 0.1 dB for all ranges
Reference impedance
[1]
2 W
[2]
, 4 W
[2]
, 8 W
[2]
, 16 W
[2]
, 50 W, 75 W, 93 W, 110 W, 124 W, 125 W, 135 W, 150 W, 250 W, 300
W, 500 W, 600 W, 800 W, 900 W, 1000 W, 1200 W, or 8000 W
Table 6-9 Supplemental measurement specifications (continued)