Technical data

6 Specifications and Characteristics
98 U3402A User’s and Service Guide
Supplemental Measurement Specifications
Display update rate
Measurement specifications
Table 6-8 Full scale display update rates
Display update rates Display counts
Slow 119,999
Medium 39,999
Fast 3,999
Table 6-9 Supplemental measurement specifications
DC voltage
Measurement method
Sigma Delta A-to-D converter
Input resistance
10 MW ± 2% range (typical)
Maximum input voltage
1000 VDC or PEAK AC on all ranges
Input protection
1000 V on all ranges
Response time
Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested
input signal at the same range.