Technical data

Operations and Features 2
U3401A User’s and Service Guide 39
Percentage (%)
This operation allows you to transfer the measurement value with a
proportional percentage display. You can set the upper limit (HI) and
lower limit (LO) to any value between 0 and ±100% of the highest range
for the present function.
The percentage operation calculates and display the measurement value
using the equation below:
Percentage (%) = [Measurement value – LO/(HI- LO)] x 100%]
When the HI limit is equal to LO limit, the percentage operation will use
the equation as below:
Percentage (%) = [(Measurement value – HI)/HI] x 100%]
Figure 2-8 Typical percentage (%) operation display
When enabled, the actual readings are shown in primary display and the
percentage’s result is shown in secondary display.
The secondary display shows OL when readings exceed the maximum
display of 999.99%.
When autoranging is enabled, this operation will be used to lock the
current range.