Technical data

6 Specifications and Characteristics
94 U3401A User’s and Service Guide
Frequency
Measurement method
Reciprocal counting technique. AC coupled input using AC voltage function
Crest factor
Maximum 3:0 at full scale
Signal level
10% of range to full scale input on all ranges
Gate time
0.1 s or 1 period of the input signal, whichever is longer.
Input impedance
1 MΩ in parallel with < 100 pF of all ranges
Maximum input voltage
1000 V RMS / 1400 V PEAK
2x10
7
V-Hz product on any range, common mode input
1x10
6
V-Hz product on any range, normal mode input
Input protection
1000 V RMS on all ranges
Response time
Approximately 1 second when the displayed reading reaches 99.9% of frequency value.
Measurement Noise Rejection
Common mode reject ratio (CMRR) for 1 kW unbalanced LO lead
50/60 Hz ± 0.1% : DC > 90 dB
Normal mode rejection ratio (NMRR)
50/60 Hz ± 0.1% : > 50 dB
Response time
Approximately 1 second when the displayed reading reaches 99.9% of frequenct value.
Table 6 - 9 Measurement specifications (continued)