Technical data

Specifications and Characteristics 6
U3401A User’s and Service Guide 93
Resistance
Measurement method
2-wire Ohms
Open-circuit voltage
Limited to +6.0 VDC
Zeroing error
0.05 Ω or less (excluding test lead resistances) in each range when Rel operation is used
Input protection
500 V DC or AC RMS
Response time
Approximately 1.5 seconds for 5 MΩ and ranges below 5 MΩ.
Diode/Continuity
Measurement method
0.83 mA ± 0.2% constant current source, open-circuit voltage limited to < 5 V
Test current
Approximately 0.5 mADC
Open-circuit voltage
Limited to < + .0 VDC
Continuity threshold
10 Ω fixed
Continuity level
Approximately below +50 m VDC
Audible tone
Continuous beep for continuity and single tone for normal forward-biased diode or
semiconductor junction.
Input protection
500 V DC or AC RMS
Table 6 - 9 Measurement specifications (continued)