Technical data

Specifications and Characteristics 6
U3401A User’s and Service Guide 91
AC voltage (True RMS AC + DC Coupling Mode)
Measurement method
AC+DC coupled true RMS — measure the AC component with up to 400 VDC bias on any
range
Crest factor
Maximum 3:0 at full scale
Input impedance
1 MΩ in parallel with <100 pF
Maximum input voltage
1000 V RMS / 1400 V PEAK
2x10
7
V-Hz product on any range, common mode input
1x10
6
V-Hz product on any range, normal mode input
Overload ranging
Will select higher range if peak input overload is detected during autoranging. Overload is
reported in manual ranging.
Input protection
1000 V RMS on all ranges
Response time
Approximate 1.5 second when the displayed reading reaches 99.9% AC RMS value of the
tested input signal at the same range.
Table 6 - 9 Measurement specifications (continued)