Technical data
6 Specifications and Characteristics
86 U3401A User’s and Service Guide
True RMS AC Current
Table 6 - 3 ACI resolution, full scale reading, and burden voltage [±(% of reading + count)]
Mode Range Resolution
Maximum
reading
Accuracy (One year; 23°C ± 5°C)
[5]
Burden
voltage
[1]
30 Hz to 50
Hz
50 Hz to 2
kHz
2 kHz to 5
kHz
5 kHz to 20
kHz
AC
coupling
500 μA
0.01 μA 510.00 1.5% + 50 0.5% + 20 1.5% + 50
3% + 75
[4]
< 0.06 V
5 mA
0.0001 mA 5.1000 1.5% + 40 0.5% + 20 1.5% + 40 3% + 60 < 0.6 V
50 mA
0.001 mA 51.000 1.5% + 40 0.5% + 20 1.5% + 40 3% + 60 < 0.08 V
500 mA
0.01 A 510.00 1.5% + 40 0.5% + 20 1.5% + 40 3% + 60 < 0.8 V
5 A 0.0001 A 5.1000
2% + 40
[3]
0.5% + 30
(< 1kHz)
— — < 0.3 V
10 A 0.001 A
20.000
[2]
2% + 40
[3]
0.5% + 30
(< 1kHz)
— — < 0.6 V
AC + DC
coupling
500 μA
0.01 μA 510.00 — 0.5% + 30 1.5% + 60
3% + 85
[4]
< 0.06 V
5 mA
0.0001 mA 5.1000 — 0.5% + 30 1.5% + 50 3% + 70 < 0.6 V
50 mA
0.001 mA 51.000 — 0.5% + 30 1.5% + 50 3% + 70 < 0.08 V
500 mA
0.01 mA 510.00 — 0.5% + 30 1.5% + 50 3% + 70 < 0.8 V
5 A 0.0001 A 5.1000 —
0.5% + 40
(< 1kHz)
— — < 0.3 V
10 A 0.001A
20.000
[2]
—
0.5% + 40
(< 1kHz)
— — < 0.6 V
Notes:
1 Typical at full range reading and voltage across the input terminals.
2 In 10 A range, > 10-20 Aac is readable for 20 seconds maximum with audio warning.
3 Input Current < 3 A RMS.
4 Input Current > 35 μA RMS.
5 Accuracy specified at input >5% of range and >1 A for 10 A range except other specified.