User`s guide

44 U2741A User’s Guide
2 Operation and Features
Triggering the U2741A
At power- up, the default trigger source is immediate. To
make a measurement, the following steps are to be followed:
Configure the U2741A by selecting the function, range,
resolution, and so on.
Specify the digital multimeter's trigger source. You can
select software bus trigger or an immediate internal
trigger (default trigger source).
Ensure that the U2741A is in wait- for- trigger state to
accept a trigger from the specified source.
Immediate triggering
In the immediate trigger mode, the trigger signal is always
present. When you place the U2741A in the wait- for-trigger
state, the trigger is issued immediately. This is the default
trigger source.
SCPI command
The following command sets the trigger source to immediate:
TRIGger:SOURce IMMediate
Software bus triggering
The bus trigger mode is initiated by sending a bus trigger
command, after selecting BUS as the trigger source.
SCPI command
The following command sets the trigger source to bus:
TRIGger:SOURce BUS
The MEASure? and READ? command overwrites the BUS
trigger and triggers the digital multimeter and returns a
measurement.
User's Guide.book Page 44 Friday, August 8, 2008 9:11 AM