Technical data
24 U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
2 Performance Test Calibration
Optional AC voltage performance verification test
These tests are not intended to be performed with every
calibration. They are provided as an aid for verifying
additional instrument specifications.
AC voltage configuration:
(CONFigure[:VOLTage]:AC [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1 Select the AC Voltage function.
2 Select each range in the order based on the input voltage
and frequency. Compare the measurement results with the
appropriate test limits shown in the Table 2- 9. (Be
certain to allow for appropriate source settling.)
[1]
Select NPLC 20 digit resolution.
Q = Quick performance verification test points
Table 2 - 9 Optional AC voltage performance verification test
Voltage Input frequency Function
[1]
Range Error from nominal 1
year
1 V 45 Hz
AC Voltage
1 V ±11mV
1 V 1 kHz 1 V ±3 mV
1 V 10 kHz 1 V ±3 mV
1 V 30 kHz 1 V ±11 mV
1 V 100 kHz 1 V ±32 mV
10 V 1 kHz 10 V ±40 mV
1 V 1 kHz 10 V ±13 mV