Technical data
20 U2741A USB Modular 5.5 Digits Digital Multimeter Service Guide
2 Performance Test Calibration
Ohms gain verification test
2- wire ohms configuration:
(CONFigure:RESistance [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
4- wire ohms configuration:
(CONFigure:FRESistance [<range|MIN|MAX|AUTO>
[,resolution|MIN|MAX|DEF>]])
1 Select the Ohms function.
2 Select each range in the order based on the input
resistance in Table 2- 5. Compare the measurement results
with the appropriate test limits shown in the following
table. (Be certain to allow for appropriate source settling.)
[1]
Select NPLC 20 digit resolution.
[2]
The specifications are for 2-wire or 4-wire ohms using the null function in
AMM software. Add 0.2 Ω error if null function is not used in AMM software.
Q = Quick performance verification test points
Table 2 - 5 Ohms gain verification test
Input Function
[1]
Range Quick check Error from nominal
1year
100 Ω
2-Wire Ohms
and
4-Wire Ohms
100 Ω ±38 mΩ
[2]
1 kΩ 1 kΩ Q ±350 mΩ
[2]
10 kΩ 10 kΩ ±35 Ω
[2]
100 kΩ 100 kΩ ±35 Ω
1 MΩ 1 MΩ ±650 Ω
10 MΩ 10 MΩ Q ±25.5 kΩ
100 MΩ 100 MΩ ±2.005 mΩ